Some products go through a final configuration and calibration before being shipped. If the calibration labs are busy this can add up to two days on to the delivery schedule but guarantees a fresh calibration.
The CR518 Acoustic Calibrator meets ANSI S1.40:2006 and IEC 60942:2003 to Class 2. It is intended for use with noise dosimeters and occupational noise meters that may require the higher calibration level.
This calibrator is included with the doseBadge Professional range of noise dosimeter kits, along with the doseBadges, charger, carrying case and software.
As soon as the calibrator is switched on it emits a 1 kHz tone at 114 dB. The calibrator should be pushed carefully (firmly, but slowly) on to the microphone.
The meter should now be adjusted accordingly. There will often be a free-field adjustment to make for the microphone in use. So, for example, you may actually be adjusting to 113.8 dB on the meter. Meters like the doseBadge Pro make these adjustments automatically and deal with any corrections needed.
Standards | IEC 60942:2003 Class 2 |
ANSI S1.40:2006 | |
Sound Level | 114.0 dB ± 0.4 dB |
Frequency | 1 kHz ± 1% |
Distortion | < 2% |
Reference conditions | 23°C, 101.3 kPa, 65% RH |
Microphone size | ½" (to IEC 1094-4) |
Battery type | 1 x 9v (PP3, 6LR61, 6F22) |
Dimensions | 121 x Ø47 mm |
Weight | 220 g with battery |
Controls | Push button start, Automatic switch off with green |
LED indication. User override of automatic switch off | |
CE classification | EN 61000-6-3:2007+A1:2011 |
EN 61000-6-1:2007 |